qCf 5 High Pressure

high pressure electrolysis

qCf Liquid Cooling High Pressure: Our system for the professional characterization of electrolysis cells for pressures up to 50 bar.
This variant of our quickCONNECT test devices was designed for PEM electrolysis at high pressures. The qCF FC5/2x63 LC 50 bar Ely was developed for operation at pressures of up to 50 bar.
  • Active electrolxsis cell area: 5 cm2
  • Operation at very high pressure: up to 50 bar
  • Liquid cooling / heating (optional equipment: HUBER ministat)

13700



GENERAL TECHNICAL DATA

Weight 10 kg
Piston diameter (actuator)
2 x Ø 62 mm; anti-rotation, low friction
Active area
5 cm2
max. operation temperature 120 °C*
Liquid cooling heating/cooling via heat bath thermostat
Dmax. back pressure 50 bar
max. contact pressure

3 kN/ 300 kg/ 675 lbf / 2.5 N/mm2

pneumatic connection (actuator)

Ø 4 mm compressed air connector

Operation type

mechanical via 5/2-way valve

Media connections (anode/cathode) Ø 6 mm connection (compatible to Swagelock)
Delivery includes qCf, power connections (Ø 6 mm/M6; Multi Contact), operating instructions
* depending on circulation thermostat and temperature control medium

Details






Optional Equipment


cF5/2x63 HP
Cell fixture with an active area of 5 cm2

HUBER ministat
Laboratory thermostat for cooling and heating

Cable set
For load connection


Anwendungen


Research & development

Quality control

balticFuelCells GmbH
Hagenower Str. 73
D-19061 Schwerin

Tel   +49 385 3993 210
info@balticfuelcells.de